1、the roads to ce marking of medical devicesfor manufacturers of devices covered by the medical device directive 93/42/eecgeneral requirementsall manufacturers of medical devices must:(custom-made devices and devices intended for clinical investigation are excluded)classify the devices (annex ix)based
2、 upon the class the device belongs to, select (article 11) and follow the appropriate conformity assessment procedure (annexes ii - vii). (see summaries for each class below.) the harmonised standard to fulfil the requirements of the quality systems in annexes ii, v and vi is en iso 13485:2003 “medi
3、cal devices quality systems requirements for regulatory purposes” where applicable.ensure that the devices fulfil the essential requirements (annex i) prepare technical documentation (annex ii, iii or vii) which shall include a general description of the product, design drawings, the results of the
4、risk analysis, a list of standards applied, test reports etc. the documentation shall make it possible to assess the conformity of the product with the requirements of the directiveestablish a system for reviewing data about products placed on the market and implement appropriate means to apply any
5、necessary corrective actionsnotify the authorities of serious accidents and narrow escapes caused by the productissue an ec declaration of conformityfile the ec declaration of conformity, the technical documentation, as well as decisions, reports and certificates from notified bodies to make them av
6、ailable to the national authorities (shall be filed for a period ending at least five years after the last product has been manufactured)all products mustcomply with the essential requirementsbe provided with ce marking, except for custom made devices and devices intended for clinical investigations
7、class i productsthe manufacturer himself declares that the products meet the provisions of the directive which apply to them, prepares a technical documentation and affix the ce mark to the device (annex vii).the manufacturer or his authorized representative for marketing of the product shall regist
8、er his activities at the competent authority of the member state where his registered business is located.exception:the device is placed on the market in a sterile condition and/or has a measuring function. the notified body shall examine the production process with regard to sterility and/or measur
9、ing function (annexes iv, v or vi).class idevice article 11:5 annex viideclaration of conformity sterile or measuring yes no *annex iv, v or viverification of notified bodycece*) limited useclass iia productsalternative 1:the notified body shall accept the manufacturers total quality system (design,
10、 manufacturing and final inspection) according to annex ii, except for section 4.alternative 2:the manufacturer himself declares that the products meet the provisions of the directive which apply to them according to annex vii. in addition the notified body examines and accepts each manufactured ite
11、m/batch according to annex iv or accepts the quality system for production and final testing according to annex v or the quality system for products according to annex vi.after having taken various administrative measures described in the relevant annex, the manufacturer then affix the ce mark and t
12、he identification number of the notified body to the product.class iiadevice article 11:2either orannex iia system with full quality assuranceannex viideclaration of conformity with annex va system with production quality assuranceannex via system with product quality assuranceannex ivec verificatio
13、n-evalutation and testing of every device or statistical verification either or or not including exam of the design (article ii, clause 3a) cececececlass iib productsalternative 1:the notified body shall accept the manufacturers total quality system (design, manufacturing and final inspection) accor
14、ding to annex ii, except for section 4.alternative 2:the notified body shall examine and assess the documentation and type-test the product according to annex iii and examine and accept each manufactured item/batch according to annex iv or accept the quality system for production and final testing a
15、ccording to annex v or the quality system for products according to annex vi.after having taken various administrative measures described in the relevant annex, the manufacturer then affix the ce mark and the identification number of the notified body to the product.class iibdevice article 11:3eithe
16、r orannex iia system with full quality assurance systemannex iiitype examination with annex via system with product quality assuranceannex va system with production quality assuranceannex ivec verification-evalutation and testing of every device or statistical verification either or or not including
17、 exam of the design (article ii, clause 3a) cececececlass iii productsalternative 1:the notified body shall accept the manufacturers total quality system (design, manufacturing and final inspection) and examine the design of the product according to annex ii.alternative 2:the notified body shall exa
18、mine and assess the documentation and type-test the product according to annex iii and examine and accept each manufactured item/batch according to annex iv or accept the quality system for production and final testing according to annex v.after having taken various administrative measures described in the relevant annex, the manufacturer then affix the ce mark and the identification number of the notified body to the product.class
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
最新文档
- 高性能湿敏传感器产品市场环境与对策分析
- 超高速电路产品市场需求分析报告
- X射线影像增强器商业机会挖掘与战略布局策略研究报告
- 直流风扇市场发展预测和趋势分析
- 铝合金游艇市场需求与消费特点分析
- 铝制日用品市场需求与消费特点分析
- 输送机保护装置商业机会挖掘与战略布局策略研究报告
- 高强瓦楞原纸产品市场需求分析报告
- 特种泵阀商业机会挖掘与战略布局策略研究报告
- 广西壮族自治区柳州市柳江区2023-2024学年七年级下学期期中考试生物卷
- 例谈数学项目学习开发路径与实践-以“量身定做的全身镜”为例
- 汽油安全技术说明书(MSDS)
- 静脉血栓栓塞症(vte)防治管理方案
- 《中国的气候-第一课时》课件
- (新插图)人教版五年级上册数学 5-2-7 实际问题与方程 用形如ax± b=c的方程解决问题 知识点梳理课件
- 医学课件:糖尿病(英文版)完整版
- 测绘投标服务方案
- 数据透视表说课
- 甘肃省托儿所幼儿园卫生保健管理实施细则
- 2023-2024学年第一学期新教科版五上科学教学计划(附进度表)
- 工程师职称工作业绩总结
评论
0/150
提交评论