1、PiCCO故障排除经验分享第1页/共16页病人情况PiCCO技术的热稀释参数在以下病人身上可能不准确:瓣膜返流,室间隔缺损主动脉瘤肺切除病人巨大肺栓塞体外循环期间严重心率紊乱严重气胸第2页/共16页病人情况PiCCO参数影响因素参数影响因素序号 疾病或治疗方法影响参数原因1心律失常SVV、PPV参数为星号SVV和PPV是30S计算一次2血温(TB)变化0.1注射冰盐水后所有参数不出由于患者自身TB变化过大,机器分不出是注射盐水后引起的TB变化,还是患者自身TB变化,无法纳入计算。3使用IABP所有连续参数由于使用IABP会对患者的压力波形造成影响,所有通过动脉压力波形测得参数都无法显示。非连续

2、参数无影响。4应用CRRT基本无影响但注意如血透管在股静脉,应不要把PiCCO导管置入同侧。同时注意TB变化。5肺大部切除、栓塞血管外肺水偏小由于肺切除或栓塞后,热稀释容积变小,血管外肺水偏小,建议参看趋势。第3页/共16页PiCCO and Dialysis 1Heise D, Faulstich M, Morer O, Brauer, Quintel MInfluence of continuous renal replacement therapy on cardiac output measurementusing thermodilution techniquesMinerva An

3、estesiol 2012; 78(3): 315-2132 ICU pts. who had either PAC (20) or PiCCO (12) andCRRT5 thermodilution injections during CRRT, 5 after CRRTstopped and 5 after CRRT restartedInjections were 20mls at 8 C F).Blood flow rate was 183 35mL/minVascular access for CRRT catheter was femoral (27),subclavian (3

4、) or internal jugular vein (2)PiCCO placed in femoral artery1第4页/共16页PiCCO and Dialysis 2The first measurements in each series after switching CRRT OFF or ON deviated the mostfrom the averageWhen these first measurements were excluded CO differed by 7%Subsequent CO measurements did not depend on CRR

5、T status.Interrupting CRRT before measuring CO is not generally recommended.However, if interrupted WAIT for the blood temperature to reach a steady state before initiatingmeasurements2第5页/共16页369 pts. on CVVH (femoral = 62 or jugular vein = 7) and PiCCO2(femoral artery)Blood pump flow rate 250mL/mi

6、n or 350mL/minFiltration flow 6000mL/hrPiCCO and Dialysis 3Dufour N, Delville M, Teboul JL, Camous L, Favier du Noyer A, Richard C, Monnet XTranspulmonary thermodilution measurements are not affected by continuousveno-venous hemofiltration at high blood pump flowIntensive Care Med 2012; 38(7): 1169

7、76第6页/共16页TDCIon was same as PCoff and TDCIoff regardless of blood flow (250 or 350 mls)No change in GEDI whether pump was on or off or set at 250 or 350mlsEVLW decreased from to in pts. with femoral dialysis catheter from TDon to TDoffMeasurements of CI, GEDI and EVLWI by PiCCO are not influenced t

8、o a clinically relevantextent by CVVH even at high pump flow rate and high filtration flow.4PiCCO and Dialysis 4第7页/共16页5VA ECMOcardiac and respiratorysupportVV ECMOrespiratory supportPiCCO and ECMOCan you use the PiCCO in patients with ECMO?ECMO has a much higher blood flow compared to RRT (6-7 l/m

9、in vs 250 mls/min)Because of this and where the catheters are placed thermodilution is not possible whilst ECMOis running as the injectate is partly lostSuggest doing TD prior to starting ECMOContinuous numbers are not effected by ECMO第8页/共16页导管位置PiCCO位置CVC位置第9页/共16页设备连接第10页/共16页操作规范第11页/共16页操作规范第12页/共16页注射液温度电缆线检查方法:导线连上机器后,往金属探针哈口气,观察机器上Tinj是否有变化T型管检查方法:观察金属弹簧在注射后是否复位压

温馨提示

  • 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
  • 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
  • 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
  • 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
  • 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
  • 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
  • 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。

评论

0/150

提交评论