1、精选优质文档-倾情为你奉上 Essential Requirements of the MDD (07/47/EC) 医疗器械指令(07/47/EC)基本要求Applicable(Y/N)适用(是/否)Requirements Standard 所符合的标准要求ManufacturersCompliance/Records 制造商符合性文件/记录Ok / Fail符合/不符合 I.GENERAL REQUIREMENTS基本要求1.The devices must be designed and manufactured in such a way that, when used under
2、the conditions and for the purposes intended, they will not compromise the clinical condition or the safety of patients, or the safety and health of users or, where applicable, other persons, provided that any risks which may be associated with their use constitute acceptable risks when weighed agai
3、nst the benefits to the patient and are compatible with a high level of protection of health and safety.This shall include: reducing, as far as possible, the risk of use error due to the ergonomic features of the device and the environment in which the device is intended to be used (design for patie
4、nt safety), and consideration of the technical knowledge, experience, education and training and where applicable the medical and physical conditions of intended users (design for lay, professional, disabled or other users).必须确保医疗器械的设计和制造在使用时或将被使用时不危害临床状况或不伤害到患者的健康,或使用者或其他人员的健康和安全,在产品使用过程中产生的与患者利益相关
5、的风险应是可接受的且符合高级别的健康安全防护措施。应包括以下内容:尽可能地降低由于医疗器械的性能特征及其使用时所处的环境(为患者安全所设计)而导致的使用错误风险。考虑技术知识、经验、教育及培训程度、使用者的医疗条件和物理条件(为非专业人士、专业人士、残疾人士及其它使用者所设计)。2.The solutions adopted by the manufacturer for the design and construction of the devices must conform to safety principles, taking account of the generally ac
6、knowledged state of the art.In selecting the most appropriate solutions, the manufacturer must apply the following principles in the following order:制造商所采用的医疗器械的设计方法和结构方法必须符合安全准则,符合大众公认的技术声明文件。制造商必须依次采用以下准则以选择最适合的方法:· eliminate or reduce risks as far as possible (inherently safe design and cons
7、truction),· 尽可能地消除或降低风险(固有安全设计和结构)· where appropriate take adequate protection measures including alarms if necessary, in relation to risks that cannot be eliminated,· 有关风险不能被消除时,应采取适当的防护措施,必要时包括安装警报装置。· inform users of the residual risks due to any shortcomings of the protection
8、 measures adopted.· 把采取了安全措施后仍然可能存在的风险告知给使用者。3.The devices must achieve the performances intended by the manufacturer and be designed, manufactured and packaged in such a way that they are suitable for one or more of the functions referred to in Article 1 (2) (a), as specified by the manufactur
9、er.制造商所设计的医疗器械必须达到预期的性能,并且其设计、生产和包装必须符合条款1(2)(a),由制造商规定的至少一项功能。4.The characteristics and performances referred to in sections 1, 2 and 3 must not be adversely affected to such a degree that the clinical condition and safety of the patients and, where applicable, of other persons are compromised duri
10、ng the lifetime of the device as indicated by the manufacturer, when the device is subjected to the stresses which can occur during normal conditions of use.涉及条款1,2和3的产品的特征和性能在制造商所指明的产品有效期内、在正常使用条件下受到挤压时必须不影响临床状况、患者安全以及其他人员的安全。5.The devices must be designed, manufactured and packed in such a way tha
11、t their characteristics and performances during their intended use will not be adversely affected during transport and storage taking account of the instructions and information provided by the manufacturer医疗器械的设计、制造和包装在其预期使用过程中其特性和性能不应受到损害,在按制造商提供的说明书和资料进行运输和储存时要保证其特性和性能的完整性。6.Any undesirable side
12、effects must constitute an acceptable risk when weighed against the performances intended.6a. Demonstration of conformity with the essential requirements must include a clinical evaluation in accordance with Annex X.当医疗器械的预期性能受到影响时,任何不良副作用必须是可接受的风险。6a.依据附录X,符合性声明和基本要求中必须含有临床评估。II.REQUIREMENTS REGARD
13、ING DESIGN AND CONSTRUCTION 设计和结构要求7. Chemical, physical and biological properties 化学性能、物理性能和生物性能7.1The devices must be designed and manufactured in such a way as to guarantee the characteristics and performances referred to in Section 1 on the "General requirements". Particular attention
14、must be paid to:医疗器械的设计和制造必须确保其符合“一般要求”中条款1所描述的特征和性能。应特别注意:· the choice of materials used, particularly as regards toxicity and, where appropriate flammability,· 所用物料的选择,尤其要注意其毒性,必要时还要考虑其可燃性。· the compatibility between the materials used and biological tissues, cells and body fluids,
15、taking account of the intended purpose of the device.· where appropriate, the results of biophysical or modelling research whose validity has been demonstrated beforehand.· 根据医疗器械的预期用途,确保其所使用的物料与生物组织、细胞和体液之间的兼容性。· 必要时,生物或模化研究结果的有效性事先应得到验证。7.2The devices must be designed, manufactured
16、and packed in such a way as to minimise the risk posed by contaminants and residues to the persons involved in the transport, storage and use of the devices and to the patients, taking account of the intended purpose of the product. Particular attention must be paid to the tissues exposed and the du
17、ration and frequency of the exposure.考虑到医疗器械的预期用途,医疗器械的设计、制造和包装必须使污染物和残余物质对在运输、存储和使用过程中所涉及到的人员和患者所造成的风险降低到最低。尤其要注意暴露组织、持续时间和暴露频率。7.3The devices must be designed and manufactured in such a way that they can be used safely with the materials, substances and gases with which they enter into contact dur
18、ing their normal use or during routine procedures; if the devices are intended to administer medicinal products they must be designed and manufactured in such a way as to be compatible with the medicinal products concerned according to the provisions and restrictions governing those products and tha
19、t their performance is maintained in accordance with the intended use.医疗器械的设计和制造必须确保其所使用的物料、物质和气体在正常使用或惯常使用情况下能安全使用。如果医疗器械用于施用医疗产品,其设计和制造必须符合医疗的相关条款和制度,并按照预期用途保持其性能。7.4Where a device incorporates, as an integral part, a substance which, if used separately, may be considered to be a medicinal product
20、 as defined in Article 1 of Directive 2001/83/EC and which is liable to act upon the body with action ancillary to that of the device, the quality, safety and usefulness of the substance must be verified by analogy with the methods specified in Annex I to Directive 2001/83/ EC.当医疗器械组合成一个整体部件,其组合所使用的
21、物质在单独使用时按照2001/83EC指令中章节1的规定可视作医疗产品,当作用于人体时,该物质在医疗器械组合部件中起辅助作用,必须根据2001/83/EC指令附录 1中规定的方法类似的方法对物质的质量、安全和效用进行验证。 For the substances referred to in the first paragraph, the notified body shall, having verified the usefulness of the substance as part of the medical device and taking account of the inte
22、nded purpose of the device, seek a scientific opinion from one of the competent authorities designated by the Member States or the European Medicines Agency (EMEA) acting particularly through its committee in accordance with Regulation (EC) No 726/2004 (1) on the quality and safety of the substance
23、including the clinical benefit/risk profile of the incorporation of the substance into the device. 关于第一段中所提到的物质,根据第726/2004号规则(1)中的规定,报告主体应对视作医疗产品组成部分的物质的效用进行验证并考虑到医疗器械的预期用途,关于物质的质量和安全,包括物质组合成医疗器械后产生的临床效益和风险,应通过其委员会寻求来自成员国或欧洲医疗代理机构指定的权威机构的科学观点。When issuing its opinion, the competent authority or the
24、 EMEA shall take into account the manufacturing process and the data related to the usefulness of incorporation of the substance into the device as determined by the notified body. 当提出观点时,权威机构或欧洲医疗代理机构应把报告主体所规定的制造过程和物质组合成医疗器械后的相关效用数据考虑在内。Where a device incorporates, as an integral part, a human bloo
25、d derivative, the notified body shall, having verified the usefulness of the substance as part of the medical device and taking into account the intended purpose of the device, seek a scientific opinion from the EMEA, acting particularly through its committee, on the quality and safety of the substa
26、nce including the clinical benefit/risk profile of the incorporation of the human blood derivative into the device. 当医疗器械组合成一个整体部件,一个人体血液衍生物,报告主体应对视作医疗产品组成部分的物质的效用进行验证并考虑到医疗器械的预期用途,关于物质的质量和安全,包括物质组合成医疗器械后产生的临床效益和风险,应通过其委员会寻求来自成员国或欧洲医疗代理机构指定的权威机构的科学观点。When issuing its opinion, the EMEA shall take int
27、o account the manufacturing process and the data related to the usefulness of incorporation of the substance into the device as determined by the notified body.当提出观点时,权威机构或欧洲医疗代理机构应把报告主体所规定的制造过程和物质组合成医疗器械后的相关效用数据考虑在内。Where changes are made to an ancillary substance incorporated in a device, in parti
28、cular related to its manufacturing process, the notified body shall be informed of the changes and shall consult the relevant medicines competent authority (i.e. the one involved in the initial consultation) , in order to confirm that the quality and safety of the ancillary substance are maintained.
29、当组合到医疗器械的辅助性物质发生变化时,特别是与之相关的制造过程发生变化时,应把这些变化通知到报告主体且咨询相关医疗权威机构(例如首次咨询的机构),以确保辅助性物质的质量和安全得到保持。The competent authority shall take into account the data related to the usefulness of incorporation of the substance into the device as determined by the notified body, in ()Regulation (EC) No 726/2004 of th
30、e European Parliament and of the Council of 31 March 2004 laying down Community procedures for the authorisation and supervision of medicinal products for human and veterinary use and establishing a European Medicines Agency (OJ L 136, 30.4.2004, p. 1) . Regulation as last amended by Regulation (EC)
31、 No 1901/2006.根据欧洲议会制定的第726/2004号(EC)规定第(*)条款和和欧洲委员会2004年3月31日制定的关于用于人类和动物的医疗产品的授权和监督程序以及建立欧洲医疗代理机构(OJL 136,30.4.2004,p.1)的决议,权威机构应把报告主体所规定的制造过程和物质组合成医疗器械后的相关效用数据考虑在内。7.5The devices must be designed and manufactured in such a way as to reduce to a minimum the risks posed by substances leaking from t
32、he device. Special attention shall be given to substances which are carcinogenic, mutagenic or toxic to reproduction, in accordance with Annex I to Council Directive 67/548/EEC of 27 June 1967 on the approximation of laws, regulations and administrative provisions relating to the classification, pac
33、kaging and labelling of dangerous substances (*) . If parts of a device (or a device itself) intended to administer and/or remove medicines, body liquids or other substances to or from the body, or devices intended for transport and storage of such body fluids or substances, contain phthalates which
34、 are classified as carcinogenic, mutagenic or toxic to reproduction, of category 1 or 2, in accordance with Annex I to Directive 67/548/EEC, these devices must be labelled on the device itself and/or on the packaging for each unit or, where appropriate, on the sales packaging as a device containing
35、phthalates.医疗器械的设计和制造必须确保由于医疗器械的物质泄漏所引起的风险降低到最小值。根据1967年6月27日制定的67/548/EEC指令附录1中的规定以及其它相关的法规、规章和管理规定中关于危险物质的分类、包装和贴标,要特别注意致癌物质、诱导有机体突变的物质或影响生殖的有毒物质。当医疗器械的部件(或医疗器械本身)用于向人体施用和/或移除来自人体的药物、体液或其它物质,或用于运输和存储这些体液或物质,包括根据67/548/EEC指令附录I中目录1或目录2被归类为致癌物质、诱导有机体突变的物质或影响生殖的有毒物质的邻苯二甲酸盐,必须在医疗器械产品上和/或在每一个包装上或必要时在其销
36、售包装上贴上标签,表明其含有邻苯二甲酸盐。If the intended use of such devices includes treatment of children or treatment of pregnant or nursing women, the manufacturer must provide a specific justification for the use of these substances with regard to compliance with the essential requirements, in particular of this p
37、aragraph, within the technical documentation and, within the instructions for use, information on residual risks for these patient groups and, if applicable, on appropriate precautionary measures当医疗器械的使用对象包括儿童或孕妇或哺乳妇女,制造商必须提供使用这些物质的详细理由及其所符合的基本要求,特别是本节要求,所要提供的资料包括技术性文件、使用说明书、残余风险相关信息,若可以,应提供相关预防措施。7
38、.6The devices must be designed and manufactured in such a way as to reduce as much as possible, risks posed by the unintentional ingress of substances into the device taking into account the device and the nature of the environment in which it is intended to be used.考虑到医疗器械及其预期使用的环境特性,医疗器械的设计和制造必须尽可
39、能的降低由于外来物质进入医疗器械所造成的风险。8.Infection and microbial contamination病毒传染和微生物污染8.1 The devices and their manufacturing processes must be designed in such a way as to eliminate or reduce as far as possible the risk of infection to the patient, user and third parties. The design must allow easy handling and,
40、 where necessary, minimise contamination of the device by the patient or vice versa during use.医疗器械及其生产过程的设计必须尽可能的消除或降低患者、使用者和第三方被传染的风险。医疗器械的设计必须易于操作,必要时,降低医疗器械在使用过程中被患者污染的程度或医疗器械对患者的污染程度。8.2Tissues of animal origin must originate from animals that have been subjected to veterinary controls and surv
41、eillance adapted to the intended use of the tissues.动物组织必须来源于组织的预期用途由兽医控制和监督的动物。Notified Bodies shall retain information on the geographical origin of the animals.认证主体应保留动物原产地相关信息。Processing, preservation, testing and handling of tissues, cells and substances of animal origin must be carried out so
42、as to provide optimal security. In particular safety with regard to viruses and other transferable agents must be addressed by implementation of validated methods of elimination or viral inactivation in the course of the manufacturing process.对动物组织、细胞和物质进行加工、存储、测试和处理时必须能够确保其最佳安全性。特别是在生产加工过程中,必须通过执行有
43、效的根除或灭菌方法对病毒和其它可传播媒介进行处理。8.3Devices delivered in a sterile state must be designed, manufactured and packed in a non-reusable pack and/or according to appropriate procedures to ensure they are sterile when placed on the market and remain sterile, under the storage and transport conditions laid down,
44、until the protective packaging is damaged or opened.以无菌形式提供的医疗器械的设计、制造和包装必须采用一次性包装,并且/或根据适当的程序确保其在规定的存储和运输条件下在上市时是无菌的并继续有效,直到其包装受损或被拆开。8.4Devices delivered in a sterile state must have been manufactured and sterilised by an appropriate, validated method.以无菌形式提供的医疗器械必须采用适当的、经认可的方法进行生产和灭菌。8.5Devices i
45、ntended to be sterilised must be manufactured in appropriately controlled (e.g. environmental) conditions.无菌医疗器械的制造必须在有适当控制的(例如:环境控制)条件中进行。8.6Packaging systems for non-sterile devices must keep the product without deterioration at the level of cleanliness stipulated and, if the devices are to be ste
46、rilised prior to use, minimise the risk of microbial contamination. The packaging system must be suitable taking account of the method of sterilisation indicated by the manufacturer.非无菌医疗器械的包装系统必须确保产品在所规定的洁净水平下没有变质,医疗器械在使用前要消毒以降低微生物污染风险。包装系统必须适当考虑制造商所指定的消毒方法。8.7The packaging and/or label of the devi
47、ce must distinguish between identical or similar products sold in both sterile and non-sterile condition.包装和/或标签必须能够区分以无菌状态或非无菌状态销售的同类产品或相似产品。9.Construction and environmental properties结构和环境特性9.1If the device is intended for use in combination with other devices or equipment, the whole combination,
48、including the connection system must be safe and must not impair the specified performance of the devices. Any restrictions on use must be indicated on the label or in the instruction for use.如果医疗器械与其它医疗器械或仪器结合使用,那么整个组合体、包括连接系统必须安全且不损害该医疗器械的特有性能。任何使用限制必须在标签或使用说明书中标明。9.2Devices must be designed and m
49、anufactured in such a way as to remove or minimise as far as possible:医疗器械的设计和制造必须尽可能消除或减少:· the risk of injury, in connection with their physical features, including the volume/pressure ratio, dimensional, and where appropriate the ergonomic features,· 由其物理特性,包括体积/压力,尺寸,必要时还包括人体工程特性所引起的受伤
50、风险。· risks connected with reasonably foreseeable environmental conditions, such as magnetic fields, external electrical influences, electrostatic discharge, pressure, temperature or variations in pressure, and acceleration,· 与一定的可预见性环境条件,例如磁场、外部电力影响,静电放电,气压,温度或压力变化和加速度有关的风险。· the risk
51、s of reciprocal interference with other devices normally used in the investigations or for the treatment given,· 与通常用于研究或治疗的其它医疗器械发生相互干扰的风险。· risks arising where maintenance or calibration are not possible (as with implants) from ageing of the materials used or loss of accuracy of any meas
52、uring or control mechanism.· 在不可能进行维修保养或校准(当移植时)时,由于所使用的物料老化或任何测量失准或操作不当所产生的风险。9.3Devices must be designed and manufactured in such a way as to minimise the risks of fire or explosion during normal use and in single fault condition. Particular attention must be paid to devices whose intended us
53、e includes exposure to flammable substances which could cause combustion.医疗器械的设计和制造必须能够降低在其正常使用过程中和出现单一故障时所产生的火灾风险或爆炸风险。应特别注意含有在使用时由于曝光而可能引起燃烧的易燃物质的医疗器械。10.Devices with a measuring function具有测量功能的医疗器械10.1Devices with a measuring function must be designed and manufactured in such a way as to provide
54、sufficient accuracy and stability within appropriate limits of accuracy and taking account of the intended purpose of the device. The limits of accuracy must be indicated by the manufacturer.具有测量功能的医疗器械的设计和制造应考虑其预期用途,并确保其在适当的精确限制值范围内具有高精确度和稳定性。制造商必须指明精确度范围。10.2The measurement, monitoring and display
55、 scale must be designed in line with ergonomic principles, taking account of the intended purpose of the device.考虑到医疗器械的预期用途,测量、监测和显示刻度的设计必须符合人类工程学原理。10.3The measurements made by devices with a measuring function must be expressed in legal units conforming to the provisions of Council Directive 80/1
56、81/EEC (OJ No L 39, 15. 2. 1980, p. 40. Directive as last amended by Directive 89/617/EEC (OJ No L 357, 7. 12. 1989, p. 28).具有测量功能的医疗器械的测量必须用 80/181/EEC 指令(OJ No L 39, 15. 2. 1980, p. 40. 由 89/617/EEC 指令(OJ No L 357, 7. 12. 1989, p. 28)修订而来)所规定的法定单位表示。11.Protection against radiation辐射防护Gen
57、eral总则Devices shall be designed and manufactured such that exposure of patients, users and other persons to radiation shall be reduced as far as possible compatible with the intended purpose, whilst not restricting the application of appropriate specified levels for therapeutic and diagnostic purposes. 在未对辐射用于治疗和诊断时的标准进行规定时,医疗器械的设计和制造应尽可能的避免患者、使用者和其他人员受到过多的辐射。Intended radiation预期辐射Where devices are designed to emit hazardous levels of radiation necessary for a specific medic
温馨提示
- 1. 本站所有资源如无特殊说明,都需要本地电脑安装OFFICE2007和PDF阅读器。图纸软件为CAD,CAXA,PROE,UG,SolidWorks等.压缩文件请下载最新的WinRAR软件解压。
- 2. 本站的文档不包含任何第三方提供的附件图纸等,如果需要附件,请联系上传者。文件的所有权益归上传用户所有。
- 3. 本站RAR压缩包中若带图纸,网页内容里面会有图纸预览,若没有图纸预览就没有图纸。
- 4. 未经权益所有人同意不得将文件中的内容挪作商业或盈利用途。
- 5. 人人文库网仅提供信息存储空间,仅对用户上传内容的表现方式做保护处理,对用户上传分享的文档内容本身不做任何修改或编辑,并不能对任何下载内容负责。
- 6. 下载文件中如有侵权或不适当内容,请与我们联系,我们立即纠正。
- 7. 本站不保证下载资源的准确性、安全性和完整性, 同时也不承担用户因使用这些下载资源对自己和他人造成任何形式的伤害或损失。
最新文档
- GB/T 41079.3-2024液态金属物理性能测定方法第3部分:黏度的测定
- 简单版工厂转让合同协议2篇
- 2024年度建筑施工沟通协调合同3篇
- 某城市地理环境课件
- 2024年度二手电动车广告投放合同2篇
- 零售店劳动合同范本
- 工程伦理课件下载
- 运维技术服务合同
- 2024年度二七区郭家咀铁三官庙N06地块安置区供排水工程合同2篇
- 公司股东入股合作协议书
- 合同到期欠款补充协议
- 本科层次职业教育装备制造类专业新形态教材建设研究
- 冬季出行安全主题班会
- 2024年学生公寓住宿协议
- 幼儿园安全守护制度
- 语文-湖南(河南)省湘豫名校联考2024年11月2025届高三上学期一轮复习诊断考试暨期中考试试题和答案
- 大学生劳动教育(2023级)学习通超星期末考试答案章节答案2024年
- 2024年中国心力衰竭诊断和治疗指南2024版
- 期中测试卷(试题)-2024-2025学年一年级上册语文统编版
- 外科手术基本操作考核评分表(切开、止血、缝合、打结与拆线)
- 国开(内蒙古)2024年《创新创业教育基础》形考任务1-3终考任务答案
评论
0/150
提交评论